Undetermined Ct Values in qPCR: What They Mean and How to Handle Them
One well in your plate comes back “Undetermined.” Before you can finish the analysis you have to answer a question the instrument software won’t answer for you: does that blank mean the gene isn’t expressed, that one reaction failed, or that something went wrong across the whole plate? The three answers lead to three completely different handling decisions, and the most common mistake — treating an undetermined Ct as a number — corrupts every fold change downstream. Here is how to read an undetermined Ct and what to do with it once you know which kind you have.
What “Undetermined” actually means
An undetermined Ct means the fluorescence never crossed the threshold within the run’s cycle limit — usually 40 cycles. It is the absence of a quantification cycle, not a Ct value of zero and not a Ct value of 40. Different instruments report the same non-event differently: QuantStudio writes “Undetermined,” Bio-Rad CFX leaves the cell empty or writes “N/A,” and some LightCycler exports put a 0 or a blank. They all mean the same thing: no detectable amplification.
First, separate the three causes
An undetermined well is only interpretable in context. Look at the rest of the plate before you decide what the blank means.
Cause 1 — a true biological negative
The gene genuinely isn’t expressed (or is below your detection limit) in that sample. The tell: the target is undetermined in a biologically sensible pattern — for example, a tissue-restricted transcript that’s blank in the wrong tissue — while your reference genes amplify normally in the same wells, and your positive control amplifies on time. If the RNA is good and the assay works elsewhere, an undetermined target is real biology, not a failure.
Cause 2 — a technical dropout in one replicate
One of three technical replicates is blank while the other two amplify cleanly around, say, Ct 26. That single dropout is almost never biology — it’s a pipetting miss, a bubble, a poorly sealed well, or cDNA stuck to the tube wall. The signature is a within-triplicate inconsistency that you’d also catch as a technical-replicate outlier: two tight values and one non-detect.
Cause 3 — a plate-wide or assay failure
Whole rows are undetermined, your reference genes are blank where they should amplify, or your positive control failed. Now the undetermined wells tell you nothing about your samples — they tell you the run failed. Common culprits are a missing master-mix component, a degraded primer stock, an inhibited reaction, or a thermal-block problem. Confirm specificity and reaction health before you trust anything on the plate; abnormal amplification-curve shapes across the plate point here, and a failed no-template control points at inhibition or contamination.
How to handle non-detects in the analysis
This is the step that’s genuinely contested, and where most spreadsheets get it wrong. Once you’ve ruled out technical failure and confirmed the non-detect is a true biological negative, you still can’t put it into a 2−ΔΔCt calculation, because there is no Ct to subtract.
Three handling strategies, in rough order of defensibility:
- Report the proportion detected, not a fabricated mean. If the target is detected in 1 of 4 biological replicates, that “1/4 detected” is the result. Averaging in an invented Ct hides the biology that matters — the gene is near or below detection in most samples.
- Treat the value as censored. Non-detects are left-censored data with a true Ct beyond the cycle limit. Methods that model this explicitly — rather than dropping or imputing — give less biased fold-change estimates when non-detects are common, as McCall and colleagues showed for qPCR differential-expression analysis (On non-detects in qPCR data, Biostatistics 2014).
- Impute a conservative bound — and say so. Some workflows substitute the cycle limit (Ct = 40) as a floor, which biases the gene toward “barely expressed” rather than “absent.” It’s defensible only if you disclose it and only when non-detects are rare; with many non-detects it distorts the comparison.
One non-detect is data; many non-detects is a design signal
If a target is undetermined in most of your samples, the honest read is usually that the assay is operating at the edge of its detection range for this material, not that expression is genuinely binary. Before reporting, consider loading more cDNA, concentrating the target, or confirming the assay’s limit of detection. A target that lands consistently at very high Ct when it does amplify is telling you the same thing from the other direction.
Summary
| What you see | Most likely cause | Handling |
|---|---|---|
| Target blank, references + control fine, sensible biology | True biological negative | Report as non-detect / proportion detected; treat as censored |
| 1 of 3 replicates blank, others tight | Technical dropout | Exclude the dropout; analyze the surviving replicates |
| Whole rows blank, references/control fail | Plate or assay failure | Don’t analyze; troubleshoot and re-run |
The decision that matters is the first one: an undetermined Ct is never a number. Diagnose which of the three causes you’re looking at, handle the technical and failure cases by exclusion or re-run, and carry true non-detects through as missing values — not zeros, not 40s. AnnealIQ parses “Undetermined,” empty, and zero-Ct cells as missing on import rather than as numbers, flags single-replicate dropouts as technical outliers, and warns when a target is missing across a group — so the non-detects in your data stay non-detects all the way to the figure.