Determining LOD and LOQ for a qPCR Assay: A Practical Method
A reviewer asks for your qPCR assay’s limit of detection, and you realize you have been reporting any Cq under 40 as a positive. That is not a limit of detection — it is a cycle cutoff. The real LOD is the lowest copy number you can reliably see, and the limit of quantification is the lowest you can trust the number for. They are different values, determined differently, and conflating them is a common way to over-claim sensitivity. This walks through both with a worked dilution series you can reproduce.
LOD and LOQ: the two limits, precisely
LOD (limit of detection) is the lowest target concentration detected with a defined probability — conventionally a 95% detection rate. At the LOD you can say “the target is present,” but you cannot put a precise number on how much.
LOQ (limit of quantification) is the lowest concentration you can measure with acceptable precision — conventionally a relative standard deviation (RSD) at or below 25% on the linear (copy-number) scale. The LOQ is always at or above the LOD; you can detect things you cannot yet quantify well.
Step 1: Build a standard curve that reaches low
Start from a quantified template (plasmid, gBlock, or quantified amplicon) and make a serial dilution that brackets the suspected limits. For LOD work you need many replicates at the low end — not the triplicate you use for the linear range. A practical layout: 6 replicates at higher levels to fix the curve, and 12 replicates at each candidate low level so a 95% rate is meaningful.
Confirm the curve’s efficiency over its quantifiable range first. Efficiency comes from the slope:
$E = 10^{-1/\text{slope}}$A slope of −3.32 gives \(E = 10^{-1/-3.32} = 10^{0.301} = 2.00\), i.e. 100% efficiency, with R² > 0.98 expected. (If that math is unfamiliar, the mechanics are in calculating qPCR efficiency from a standard curve.) With slope −3.32 and an intercept of 40, the predicted Cq values are:
| Copies / reaction | log₁₀(copies) | Predicted Cq |
|---|---|---|
| 1000 | 3.00 | 30.04 |
| 100 | 2.00 | 33.36 |
| 10 | 1.00 | 36.68 |
| 5 | 0.70 | 37.68 |
| 1 | 0.00 | 40.00 |
Step 2: Find the LOD by counting detections
Run the low-concentration replicates and count how many amplify (any reportable Cq), not the Cq value itself. Detection is binary here. Suppose 12 replicates at each level give:
| Copies / reaction | Detected / 12 | Detection rate |
|---|---|---|
| 10 | 12 / 12 | 100% |
| 5 | 12 / 12 | 100% |
| 3 | 11 / 12 | 91.7% |
| 2 | 9 / 12 | 75% |
| 1 | 5 / 12 | 41.7% |
The LOD is the lowest level meeting the 95% threshold. Here 5 copies is detected 100% of the time but 3 copies only 91.7% — below 95%. So LOD = 5 copies/reaction. Below that, detection becomes a coin-flip and Poisson sampling of template at single-digit copy numbers dominates.
Step 3: Find the LOQ by checking precision
Detection is not enough for quantification. At each candidate level, back-calculate copy number per replicate from the standard curve, then compute the RSD:
$\text{RSD} = \frac{s}{\bar{x}} \times 100\%$At 10 copies/reaction, six back-calculated values come out as 11.5, 9.2, 10.9, 8.4, 11.0, 9.0 copies. The mean is 10.0 and the sample standard deviation is 1.29, giving:
At 5 copies/reaction, the six values are 6.8, 3.1, 5.9, 2.4, 7.2, 4.1. Mean = 4.92, SD = 2.00, RSD = 2.00 / 4.92 × 100% = 40.7% — well above 25%. So 5 copies is detectable but not quantifiable.
The lowest level meeting the ≤25% RSD criterion is 10 copies. So LOQ = 10 copies/reaction — double the LOD, which is the expected relationship: you can detect 5 copies every time but the number attached to it scatters too much to report.
Step 4: Sanity-check and report
Confirm the result is reasonable: LOQ (10) ≥ LOD (5), both fall inside the standard-curve range, and the curve efficiency over the quantifiable range is 90–110%. Report both limits with the method used to derive them — “LOD = 5 copies/reaction (95% detection, n=12); LOQ = 10 copies/reaction (RSD ≤ 25%, n=6)” — not a bare Cq cutoff.
Where this connects
LOD/LOQ work assumes your Cq values are themselves well-defined, which depends on consistent threshold and baseline settings across the plate — a shifted threshold moves every Cq and drags the limits with it. And because copy-number limits come from a standard curve, this is fundamentally absolute quantification territory; relative-expression studies do not report LOD/LOQ in copies, though the detection-reliability idea still bounds how low a Cq you should trust.
The canonical reference is Forootan et al., “Methods to determine limit of detection and limit of quantification in quantitative real-time PCR” (Biomolecular Detection and Quantification, 2017), and the MIQE guidelines (Bustin et al., Clinical Chemistry 2009) specify reporting LOD as part of assay validation.